FOXBORO is now a brand of Schneider in France, which mainly manufactures instruments for industrial control and process automation. Schneider Electric (French: Schneider Electric) was founded in 1836, is a large energy management company, one of the optimization solutions provider, the main products include circuit breakers, sensors, controllers, etc., for the countries energy facilities, industry, data centers and networks, buildings, to provide total solutions. Which in the energy, infrastructure, industrial process control, building automation, data centers and networks and other markets.
FOXBORO main products: FOXBORO sensors, FOXBORO circuit breakers, FOXBORO control valves, FOXBORO actuators, FOXBORO converters, FOXBORO relays, FOXBORO controllers, FOXBORO recorders and indicators, FOXBORO thermocouple bushing, FOXBORO DEWCEL Dew point meters, FOXBORO filter regulators, FOXBORO pneumatic transmitters and more.
Descrption of FOXBORO Analyzer 873EC-AIPFGZ-7
873EC Series Electrochemical Analyzers for Electrodeless Conductivity Measurement.The 873EC Analyzer measures the conductivity of aqueous solutions. Its measurement display may be read in either μS/cm, mS/cm, or percent (%). Solution temperature is also measured by the 873EC for automatic temperature compensation and may be displayed whenever the user wants. It provides an isolated output signal proportional to the measurement for transmission to an external receiver. The plastic panel-mounted analyzer transmits one output signal only, while the metal enclosure analyzers can transmit two output signals.
Supply Voltages
–A 120 V ac
–B 220 V ac
–C 240 V ac
–E 24 V ac
–J 100 V ac
Supply Frequency
50 or 60, ±3 Hz
Output Signal
4 to 20 mA isolated
0 to 10 V dc isolated
0 to 20 mA isolated
Ambient Temperature Limits
–25 to +55°C (–13 to +131°F)
1. Introduction MI 611-167 – February 2016
17
Measurement Ranges
50, 100, 200, and 500 μS/cm; 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, and 2000 mS/cm;
depends on sensor used. Chemical concentration ranges are available for several electrolytes:
HCl, 0-15%
NaCl, 0-25%
H3PO4, 0-35%
HNO3, 0-10%
H2SO4, 0-25%
H2SO4, 99.5-93% at 50 °C (122°F)
Oleum, 42-18% at 65 °C (149°F)
NaOH, 0-15% at 25 °C (77°F)
NaOH, 0-20% at 100 °C (212°F)
Temperature Measurement Range
–17 to +199°C (0 to 390°F) w/100 Ω RTD
–17 to +121°C (0 to 250°F) w/100 kΩ thermistor
Temperature Compensation Range
See Table 7.
Relative Humidity Limits
5 to 95%, noncondensing
Accuracy of Analyzer
±0.5% of upper range limit
Analyzer Identification
Refer to Figure 2.
Dimensions
Plastic:92(H) x 92(W) x 183(L) mm [3.6 (H) x 3.6 (W) x 7.2 (L) inches]
Metal: 92(H) x 92(W) x 259(L) mm [3.6 (H) x 3.6 (W) x 10.2 (L) inches]
Enclosure/Mounting Options
–P Plastic /Panel Mount
–W Metal/Panel Mount
–X Metal/Surface Mount
–Y Metal/Pipe Mount
–Z Metal/Movable Surface Mount
FOXBORO 873EC-AIPFGZ-5
FOXBORO 873EC-AIXCG
FOXBORO 873EC-AIYFGZ
foxboro 873ec-biyfgz-7
FOXBORO 873EC-AIYFGZ-7
FOXBORO 873EC-AIPFG2