Midas® Sensor Cartridge Specifications Sulfur Dioxide (SO2) MIDAS-E-SO2, MIDAS-S-SO2
Gas Measured Sulfur Dioxide (SO2)
Cartridge Part Number MIDAS-S-SO2 1 year standard warranty,
MIDAS-E-SO2 2 year extended warranty
Sensor Technology 3 electrode electrochemical cell
Measuring Range SO2 0 – 8ppm
Minimum Alarm 1 Set Point 0.95ppm
Lower Detection Limit 0.70ppm
Linearity <± 10% of measured value
Repeatability <± 2% of measured value
Resolution 0.05ppm
Response Time t62.5 ≤ 5 seconds
Sensor Cartridge Life
Expectancy
≥ 24 months under typical application conditions
Operating Temperature
Effect of Temperature
Zero
Sensitivity
0°C to +40°C (32°F to 104°F)
<± 0.03ppm / °C
<± 0.4% of measured value / °C
Operating Humidity
Effect of Humidity
Zero
Sensitivity
15 to 90% RH
No effect
<± 1% of measured value / % RH
Operating Pressure 90 – 110kPa
Effect of Position No effect in typical application
Long Term Drift
Zero
Sensitivity
TBA
<± 2% of measured value / year
Calibration Gas Sulfur Dioxide (SO2)
Bump Test Gas Sulfur Dioxide (SO2)
Warm Up Time < 10 minutes
Storage Temperature +5°C to +25°C (+41°F to +77°F)
MIDAS-E-SO2 Honeywell MIDAS Sensor Cartridges