Midas® Sensor Cartridge Carbonyl Sulfide (COS) MIDAS-X-COS
Each Midas® sensor is potentially cross sensitive to other gases and this may cause a gas reading when exposed to other gases than those originally designated. The table below presents typical readings that will be observed when a new sensor cartridge is exposed to the cross sensitive gas (or a mixture of gases containing the cross sensitive species).
Gas Measure Carbonyl Sulfide (COS)
Cartridge Part Number MIDAS-X-COS 2 year extended warranty
Sensor Technology 4 electrode electrochemical cell
Measuring range (ppm) COS 0-100 ppm
Minimum Alarm 1 Set Point 10 ppm
Repeatability < ± 30 % of full scale
Linearity < ± 30 % of full scale
Response time t62.5 ≤ 60 seconds
Sensor Cartridge Life Expectancy ≥ 24 months under typical application conditions
Operating temperature
Effect of Temperature
Zero
Sensitivity
0 °C to + 40 °C (32 °F to 104 °F)
< ± 10 ppm (at 10 °C to 30 °C)
< ± 30 % of full scale (at 10 °C to 30 °C)
Operating humidity
Effect of Humidity
Zero
Sensitivity
15-90 %RH (Sensor Datasheet)
< ± 10 ppm (at 45 %RH)
< ± 30 % of full scale (at 45 %RH)
Operating Pressure 90 – 110kPa (Sensor Datasheet)
Effect of Position No effect in typical application
Long Term Drift
Zero
Sensitivity
< ± 10 ppm
< ± 30 % of full scale
Calibration Gas Carbonyl Sulfide (50 ppm COS)
Challenge Gas (Bump Test) Carbonyl Sulfide (50 ppm COS)
Warm Up Time ≤ 60 minutes
Storage Temperature +3 °C to +20 °C (+37 °F to +68 °F) (Sensor Datasheet)
Honeywell Analytics MIDAS Sensor Cartridge, Carbonyl Sulfide (COS) 0-100ppm, 2 years - MIDAS-X-COS
MIDAS-X-COS;MIDAS-E-CO2;MIDAS-L-O2S
MIDAS-LEX;MIDAS-E-CFX;MIDAS-E-TEO
MIDAS-E-SHX;MIDAS-E-HCH;MIDAS-E-HCN
MIDAS-E-PH3;MIDAS-E-SO2;MIDAS-E-CFH
MIDAS-E-BR2;MIDAS-I-N20;MIDAS-E-SHL
MIDAS-E-NH3;MIDAS-E-H2S;MIDAS-E-NO2
MIDAS-E-03X;MIDAS-E-COH;MIDAS-X-IPA
MIDAS-E-HFX;MIDAS-E-HCL;MIDAS-E-HAL
MIDAS-E-03H;MIDAS-I-SF6;MIDAS-E-NOX
MIDAS-E-PHX;MIDAS-E-LEO;MIDAS-E-COX
MIDAS-E-B2H;MIDAS-E-LEL;MIDAS-E-LEB
MIDAS-X-IPA;MIDAS-E-XHF;MIDAS-I-CO2